Electrostatic Field Meter Verification Kit
Hisco #:7100-EFM51-VK-52162
MFG #:7100-EFM51-VK
Item must be ordered in multiples of 1
Quantity | Price | Save |
---|---|---|
1 | 4375.23 | |
3 | 4162.13 | Save 5% |
The Transforming Technologies 7100-EFM51-VK EFM51.VK Verification Kit is a full-featured Static Field Meter Kit designed to test static fields, verify ionizers, and assess the performance of ESD floors in combination with a person, per ANSI/ESD STM 97.2 and IEC Standard 61340-4-5.
- Field Meter
The EFM51 measures static voltages within ±20kV (20,000V) using a microcontroller that calculates the field strength (V/m) at pre-selected distances (1cm, 5cm, 10cm, and 20cm). Results are displayed on a 12-digit alphanumeric LCD screen, with two LED guide ring lights to help position the field meter at the correct distance from a charged test object. - Charge Plate System
The EFM51.VK evaluates ESD ionizer efficiency and performs static surveys, assessing charge accumulation in electronic manufacturing and industrial operations. It measures the decay time from +1,000V to +100V or -1,000V to -100V with an integrated start-stop timer. The plate is charged manually using a handheld high voltage generator. - Walking Test
This feature measures the electrostatic charge of a person walking on an ESD-safe floor with ESD-safe footwear. It provides data on the actual charge on a person while walking, which is used to determine the effectiveness of the floor/footwear grounding system. The EFM51 holds the PEAK value of the measured body voltage, as well as the LIVE static voltage on the operator’s body.
The EFM51 meets periodic test requirements per Compliance Verification ESD TR53 and conforms to ANSI/EOS/ESD (S4.1, S7.1, S12.1, S2.1).
Product Description
The Transforming Technologies 7100-EFM51-VK EFM51.VK Verification Kit is a full-featured Static Field Meter Kit designed to test static fields, verify ionizers, and assess the performance of ESD floors in combination with a person, per ANSI/ESD STM 97.2 and IEC Standard 61340-4-5.
- Field Meter
The EFM51 measures static voltages within ±20kV (20,000V) using a microcontroller that calculates the field strength (V/m) at pre-selected distances (1cm, 5cm, 10cm, and 20cm). Results are displayed on a 12-digit alphanumeric LCD screen, with two LED guide ring lights to help position the field meter at the correct distance from a charged test object. - Charge Plate System
The EFM51.VK evaluates ESD ionizer efficiency and performs static surveys, assessing charge accumulation in electronic manufacturing and industrial operations. It measures the decay time from +1,000V to +100V or -1,000V to -100V with an integrated start-stop timer. The plate is charged manually using a handheld high voltage generator. - Walking Test
This feature measures the electrostatic charge of a person walking on an ESD-safe floor with ESD-safe footwear. It provides data on the actual charge on a person while walking, which is used to determine the effectiveness of the floor/footwear grounding system. The EFM51 holds the PEAK value of the measured body voltage, as well as the LIVE static voltage on the operator’s body.
The EFM51 meets periodic test requirements per Compliance Verification ESD TR53 and conforms to ANSI/EOS/ESD (S4.1, S7.1, S12.1, S2.1).
Technical Information
Item must be ordered in multiples of 1